● Remote monitoring
● LED indicators:On, Insulation, Overload, Overtemp
● Mute button
● Test button
● Auxiliary power supply: DC24V
● RS485(MODBUS-RTU)
Overview
AID series alarm and display indicator are remote display and audible and visible alarm devices. AID duplicates fault, alarm and operating message of monitoring devices. When fault occurs, AID gives audible and visible signal to medical staff and technical staff. Additionally, response value of AIM-M series monitoring device can be set by AID remotely. AID Series alarm and display device have three types: AID10, AID120 and AID150, which shall be arranged in operating room or nursing station so that it can be permanently monitored by medical staff.Dimension
Installation
Typical Connection
Model Selection
Application programme-ICU/CCU
Typical Connection
Model | Selection Description |
AID120 | It can monitor one set of AIM-M100 insulation monitor and be used for installation by embedding into wall. It can be applied to monitor operating room or ICU only using single set of isolated power. |
AID150 | It can monitor 16 sets of AIM-M100 insulation monitoring instruments to the maximum and AIM-R100 residual current monitor which can be used for installation by embedding into wall. It is fit for the centralized monitoring on operating room or ICU or other medical sites. |