● Insulation monitoring for medical IT systems
● Load and temperature monitoring of isolation transformer
● Device disconnection monitoring
Overview
AIM-M10 medical insulation device instrument adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit. AIM-M10 is usually combined with AITR series medical isolation transformer, AKH-0.66P26 current transformer and AID10/150 centralized alarm display device.Dimension
Wiring
Model Selection
Application programme-ICU/CCU
Typical Connection
Auxiliary power supply | Voltage | AC220V (±10%) | Temperature measurement | Thermistor | Pt100 |
Frequency | 50/60Hz | Measuring range | -50~+200℃ | ||
Power consumption | <5W | Alarm value range | 0~+200℃ | ||
Insulation monitoring | Measuring range | 10~999kΩ | Alarm output | Output mode | 1 relay output |
Absolute percentage | 0~±10% | Contact capacity | AC 250V/3ADC 30V/3A | ||
Alarm value | 50~999kΩ | Environment | Operating temperature | -10~+55℃ | |
Response time | <2s | Storage temperature | -20~+70℃ | ||
Measuring voltage | <12V | Relative humidity | 5~95%, non-condensate | ||
Measuring current | <42μA | Altitude | ≤2500m | ||
Load current | Measuring range | 2.1~50A | Communication | RS485, Modbus-RTU | |
Alarm value | 5~50A | Rated impulse voltage/pollution degree | 4KV/Ⅲ | ||
Measuring accuracy | ≤±5% | EMC/EMR | Conform to IEC 61326-2-4 |