Medical IT system insulation monitoring device and fault location system is suitable for hospital operating room, ICU (CCU) and other important places, and can provide a safe place for such continuous, reliable power solutions.
1)Primary connection diagram and field distribution display
2)Real-time data acquisition and display
3)Fault Alarm
4)Remote parameter setting and inquiry
5)Graphic display
Display the insulation status, load status, and the isolation transformer temperature of each set of isolated power systems in graph form.
● Insulation monitoring for medical IT systems
● Load and temperature monitoring of isolation transformer
● Device disconnection monitoring
● LED indicators: On, Insulation, Overload, Overtemp
● Alarm relay output
● Test button
● Insulation fault location
● RS485 (MODBUS-RTU) + CAN
AID150 Alarm and Display Device
● Remote monitoring
● LED indicators:On, Insulation, Overload, Overtemp
● Mute button
● Test button
● Auxiliary power supply: DC24V
● RS485(MODBUS-RTU)
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Quisque placerat quis nulla eu faucibus. Donec ac odio quis libero commodo ornare. Nam tellus tortor, tempus et consectetur at, convallis nec justo.
Curabitur ornare vel ante a elementum. Suspendisse sagittis nisl nisl, vulputate gravida leo elementum eu.
● Generating Fault Location Signal
● Detection of L1 and L2 disconnection
● LED indicators: On, Comm, Test
● CAN bus
● Auxiliary power supply: DC24V
● Auxiliary power supply to AID series
● AC 220V input, DC 24V output
● Completely Isolated Linear Transformer
● Power-on indicator
● Insulation fault location in AC IT systems
● AIL150-4: 4 measuring channels
● AIL150-8: 8 measuring channels
● LED indicators: On, Comm
● CAN bus
● Auxiliary power supply: DC24V
● Isolating transformers for medical IT systems
● Double insulation treatment, Electrostatic shielding layer
● PT100 temperature sensor
● Rated power: 3.15…10KVA
● CE
AKH-0.66P26 Protective Current Sensor
● Matching with AIM-M Series Insulation Monitor
● Maximum Current: 50A
● Transformation ratio: 2000:1